Vspw, Vspw vertical sync pulse width – Quantum Data 802BT User Manual

Page 577

Advertising
background image

Video Test Generator User Guide

567

VSPW

Vertical Sync Pulse Width

Class

Format parameter setting - Synchronization

Description

Establishes the width of the vertical sync pulse in lines. If the type specified for the

selected sync signal (see

SSST

,

ASCT

,

DSCT

, or

DSST

commands) is one of the CCIR

types, then the actual sync pulse width output by the generator will be 1/2 line shorter than

the whole number specified. The VSPW? query returns the current setting of VSPW.

Command syntax

VSPW lines

Query syntax

VSPW?

Query response

lines

Other required

commands

The

FMTU

command instructs the generator to use the new setting. The

ALLU

command

updates hardware to the new setting and redraws the test image.

Example

VSPW 3 // Set V sync width to 3 lines in buffer
FMTU

// Update hardware to current buffer contents

Advertising
This manual is related to the following products: