Appendix A Command Reference

PSPM

Probe Sync Pulse Mode

Class

Special Sync Probe Pulse

Description

Determines what lines in the frame or field the probe pulse occurs on the special sync

output. The PSPM? query returns the current PSPM value.

Command syntax

PSPM mode

mode integer

0 = once per frame

1 = every active line in frame

3 = every line in frame

Query syntax

PSPM?

Query response

0, 1, or 3

Other required

commands

The

FMTU

command instructs the generator to use the new settings. The

ALLU

command

updates the hardware to the new settings and redraws the test image.

Related commands

PSPG

,

PSPP

,

PSPW

,

PSHD

,

PSVD

Example

PSPG 1

// Enables probe pulse on special sync

PSPP 1

// Sets probe pulse polarity to positive

PSPW 10

// Sets probe pulse width to 10 pixels

PSHD 200

// Sets probe pulse horizontal delay to 200 pixels

PSVD 300

// Sets probe pulse vertical delay to 300 pixels

PSPM 1

// Sets probe pulse to occur on each active line in a frame

ALLU

// Applies the settings