Calibrating the generator

The 802BT/R generator can calibrate itself. Once calibrated, the generator does not

require periodic calibration. There are no physical controls to adjust. All calibration is

electronic. Calibration factors are saved in non-volatile memory.

The calibration accuracy of the generator for analog video voltage swing is specified to be

±14mV (or ±2% for a nominal 700mV signal). Using a typical oscilloscope such as a

Tektronix VM700T, which might have a swing measurement accuracy of ±14mV, we

should expect readings between 674mV to 726mV.

You can also set the analog output calibration factors to values other than those set by the

generator's own self-calibration function. Each channel is calibrated using two

independent zero calibration points, and two independent slope calibration points.

Interpolation equations for the different video types are used to set the caldac values that

controls the output sync and voltage levels to the correct values.

Follow the procedures below to calibrate the generator.

To self-calibrate the generator using the front panel:

1. It is not necessary to load the outputs of a generator during self-calibration. However,

attaching a nominal load may yield a slight increase in accuracy.

2. Simultaneously depress all three R, G and B video gate keys and power cycle the

generator.

The LCD will display a special message indicating that it is about to do a

self-calibration.

3. Release the R, G and B video gate keys and the generator will begin its self-calibration

routine.

You will see calibration factors flash on the display. Observe the display for any error

messages that might occur. If an error message appears, make a note of it and call your

technical support representative for further assistance.

To self-calibrate the generator using the command interface:

1. Establish a terminal session with the generator. See “

Setting up a terminal connection

with the generator

” on page 14.

2. Enter the following command:

SCAL

The generator then goes through a complete calibration procedure.

To view analog calibration factors:

1. Enter the following commands to view the analog video Red, Green, Blue calibration

factors: