Calibrating frequency – Quantum Data 802BT User Manual

Page 58

Advertising
background image

48

Chapter 3

Administrative Tasks

CALF

?

The generator presents the following in response:

1089, 2589, 1993, 3223, 1148, 2633, 2059, 3279, 1000, 2505, 1990,
3219, 793, 787, 3382, 3509

Calibrating frequency

All frequencies associated with the generator are derived from a common frequency

reference, which can be calibrated using the rate calibration command

RATC

.

To reduce all frequency errors to zero:

1. Measure any convenient frequency from the generator using a very accurate frequency

counter.

For example, measure the horizontal TTL separate sync frequency with the DMT0660

format loaded and the horizontal rate temporarily set to 100KHz. Enter the following:

FMTL

DMT0660; HRAT 100E3;

ALLU

2. Measure the TTL horizontal sync frequency with the frequency counter.

3. Divide the frequency that you expect by the frequency that you measure on the

frequency counter and enter that ratio using the

RATC

command.

For example, if you measure 99.9955782499875 kHz, when you expect

100.000000000KHz, then divide 100.000000000KHz by 99.9950002499875kHz and

enter the following on the command line:

RATC

1.0000442194553987

This will raise all frequencies output by the generator and make them exact.