802BT/802R Video Test Generator User Guide

723

3006

Nothing to save

An attempt was made to save the contents of a custom image editing buffer when there

was no open custom image editing session to save.

3007

Overwrite ROM Image

A user defined test image can not be saved using the same name as that of a built-in test

image (case insensitive).

3008

Image save failed

Saving the contents of the custom image editing buffer failed for a reason other than those

reported by error codes 3005 or 3006.

3010

R, G or B 8-Bit video DAC setting out of range

A combination of system calibration factors, user calibration multiplier settings and video

output level settings have resulted in one or more calculated input values to go below zero

or above the analog video Digital-to-Analog Converter’s maximum limit of 255.

3011

R, G or B 6-Bit video DAC setting out of range

A combination of system calibration factors, user calibration multiplier settings and video

output level settings have resulted in one or more calculated input values to go below zero

or above the analog video Digital-to-Analog Converter’s maximum limit of 63.

3015

R, G or B 10-Bit video DAC setting out of range

A combination of system calibration factors, user calibration multiplier settings and video

output level settings have resulted in one or more calculated input values to go below zero

or above the analog video Digital-to-Analog Converter’s maximum limit of 1023.

3020

Kill EPROM image

Built in test images can not be removed from the firmware.

3025

Image not found

A test image of a given name could not be found as either a built-in test image in the

firmware or as a user defined custom image.

3050

Grayscale video DAC setting out of range

One or more internal calculations have resulted in an analog video DAC input grayscale

setting below zero or greater than 100% video level.