Quantum Data 802BT User Manual

Page 56

Advertising
background image

46

Chapter 3

Administrative Tasks

To re-initialize generator through system reset:

1. Hold down ACS and DSS while powering up the generator.

A confirmation message is displayed.

2. Press the Outputs key to continue, or press the Step key to cancel the operation.

To re-initialize and calibrate the generator using command interface:

1. Establish a terminal session with the generator (see page 14).

2. Enter the following command:

INIT

The generator then goes through a complete self-test and self-calibration procedure.

To re-initialize without calibrating the generator using the command line:

1. Establish a terminal session with the generator (see page 14).

2. Enter the following command:

BOIT

The generator is re-intialized.