Step Size and Offset Menu

The Step Size and Offset Menu lets you set the center frequency step size and frequency offset
parameters.

Table 4-4.

Step Size and Offset Menu

CF Step Size: To set the center frequency step size, press the CF Step Size button and
select the desired step size from the fly-out menu.

To set the span step size manually, select Manual in the fly-out menu, then press in the
CF Step Size parameter field and enter a value using the numeric keys. When a new
value is entered, the terminator fly-out will appear with the available units. Press on the
desired units to terminate the entry.

Range: 1/10 Span, ½ Span, %Span, =Center, =Marker, =Span, Manual
Resolution: 1 Hz (for Manual), 0.01 Hz or 1 Hz (for %Span)
Default: 1/10 Span
Terminators: GHz, MHz, kHz, Hz, Cancel
(Applies only to choices where a numeric value is entered)

Frequency Offset: To set the frequency offset, press in the Frequency Offset parameter
field and enter a new value using the numeric keys. When a new value is entered, the
terminator fly-out menu will appear with the available units. Press on the desired units to
terminate the entry.

Range: –100 GHz to +100 GHz
Resolution: 1 Hz
Default: 0 Hz
Terminators: GHz, MHz, kHz, Hz, Cancel