Anritsu Series MS278XB User Manual

Page 244

Advertising
background image

Operation Verification

Measurements

5-88

MS278XB OM

9. Adjust the MG369XB RF output so that the power meter input B reading is –10 dBm. Record the

corresponding power meter input A reading in Table 5-10.

10. Repeat steps 7 and 8 for the other frequencies listed in Table 5-10.
11. Disconnect the Phase Matched N-male to N-female adapter from the 10 dB attenuator, and then

connect a N-male to N-male Phase Matched adapter between the 10 dB pad and the MS2781B RF
input.

12. Set up the MS2781B as follows:

a. Center Frequency: 50 MHz
b. Span: 300 kHz
c. RBW:50 kHz
d. VBW: 500 Hz
e. Sweep Time: 100 ms
f. Attenuation: 10 dB
g. Reference Level: 0 dBm

13. Set the frequency on the MG369XB to 50 MHz, then set the output power level so that the sensor A

reading matches the corresponding value as recorded in column 2 of Table 5-10.

14. Set the Marker to Peak. Record the measured Marker value in Table 5-10.
15. Repeat steps 11 and 12. Record the measured results in Table 5-10. Verify that the measured

results are within specifications.

Note: Set the Cal Factor on the power meter to match the frequency being measured.

Table 5-10. Frequency Response Test

Frequency

Sensor A Power Reading for

–10 dBm Output at Adapter

MS2781B Marker Reading

Specifications

–10 dBm

50 MHz

±0.4 dB

500 MHz

±0.4 dB

1000 MHz

±0.4 dB

1500 MHz

±0.4 dB

2000 MHz

±0.4 dB

3000 MHz

±0.4 dB

4000 MHz

±0.4 dB

5000 MHz

±0.4 dB

6000 MHz

±0.4 dB

7000 MHz

±0.4 dB

7800 MHz

±0.4 dB