Operation Verification

Measurements

5-92

MS278XB OM

13. Record the new marker level readout value on the display to the corresponding cell in Table 5-12.
14. Calculate the amplitude deviation by subtracting the new marker value from the 30 kHz RBW

marker reading.

15. Verify that the amplitude deviation is

<

0.1 dB.

16. Repeat steps 11 through 16 for the rest of RBW and SPAN combinations in Table 5-12.

Table 5-12. Resolution Bandwidth Switch Uncertainty Test

RBW

SPAN

MS2781B Marker

Reading

Amplitude Deviation

from 30 kHz RBW (dB)

Specifications

10 Hz

10 kHz

0.15 dB

100 Hz

1 kHz

0.15 dB

300 Hz

5 kHz

0.15 dB

3 kHz

30 kHz

0.15 dB

10 kHz

100 kHz

0.15 dB

30 kHz

500 kHz

0.15 dB

100 kHz

1 MHz

0.15 dB

300 kHz

5 MHz

0.15 dB

1 MHz

10 MHz

0.15 dB

3 MHz

50 MHz

0.15 dB