VSA Modulation Parameters Setup Dialog

The VSA Modulation Parameters dialog allows you to specify the modulation setting for your measurement.

Modulation Tab

Modulation Settings:
Format:

BPSK: Binary Phase Shift Keying
QPSK: Quadrature Phase Shift Keying (default)
Pi4QPSK: Pi/4 Quadrature Phase Shift Keying
8PSK: Octet Phase Shift Keying
3Pi8PSK: 3Pi/8 Phase Shift Keying
16QAM: 16 Value Quadrature Amplitude Modulation
32QAM: 32 Value Quadrature Amplitude Modulation
64QAM: 64 Value Quadrature Amplitude Modulation
128QAM: 128 Value Quadrature Amplitude Modulation
256QAM: 256 Value Quadrature Amplitude Modulation

Symbol Rate: Sets the symbol rate for the
measurement.

Range: 10 kHz to 20 MHz
Default: 3.84 MHz.

Capture Time: Sets the length of the time during which
samples are captured.

Range: 1s to 5 µs*
Terminators: us, ms, s, ks, Cancel

* Constrained by the following:
10,000

(Symbol Rate x Capture Time

100

Filtering:
Filter Type: Selects from the following receiver filters:

Low Pass
Nyquist
Root Nyquist (default)

Rolloff Factor: Sets the roll-off factor for the
measurement.

Range: 0.1 to 1.0 in steps of 0.01
Default: 0.22.

Figure 4-19. VSA Modulation Parameters Dialog