Anritsu Series MS278XB User Manual

Page 232

Advertising
background image

Operation Verification

Measurements

5-76

MS278XB OM

9. Turn on Marker and perform Marker Peak search. Read the marker frequency and verify that the

value is within specifications using the formula below:

± ((marker freq x reference accuracy) + span accuracy +
(0.05 x RBW) + (0.5 x last digit))

For technical specifications, refer to the Signature Technical Data Sheet, part number: 11410-
00396, located in Appendix A of the Signature operation manual.

10. Repeat steps 8 through 9 for other RBW/Span combinations listed in Table 5-4.
11. Repeat steps 7 through 10 for 7900 MHz.

Table 5-4.

Frequency Readout Accuracy Test

FREQ

(MHz)

RBW

(Hz)

Span

(Hz)

Specification

(Hz)

Marker Readout

Frequency

100

100

100

100

1000

100

10,000

1000

1,000

1000

10,000

1000

100,000

10,000

10,000

10,000

100,000

10,000

1,000,000

7900

100

100

100

1,000

100

10,000

1000

1000

1000

10,000

1000

100,000

10,000

10,000

10,000

100,000

10,000

1,000,000

Advertising