Trigger Menu

The Trigger menu lets you configure the following:

Table 4-18. Trigger Menu

Trigger Source: To change the trigger source, press on the Trigger Source button and
select the trigger source from the fly-out menu selections listed under Trigger Level
below:

Free Run: No Settings

Video:
Range: Reference Level to (Reference Level – 10 * Scale/Div)
Resolution: 0.1 dB
Default: Reference Level – 0.5 * (10 * Scale/Div)
The following unit terminators are supported when the appropriate amplitude units
are selected: dBm, dBmV, dBµV, W, mW, µW, nW, pW, f W , aW, V, mV, µV, nV, pV

IF Power (Wide BW):
Range: (–10 dBm to –50 dBm) + Attenuator Value
Resolution: 1 dB
Default: –10 dBm + Attenuator Value
The following unit terminators are supported when the appropriate amplitude units
are selected: dBm, dBmV, dBµV, W, mW, µW, nW, pW, f W , aW, V, mV, µV, nV, pV

External TTL: Level = 1.4 V (cannot be changed)

External:
Range: –10V to 10V (External)
Resolution: 100 mV
Default: TTL (1.4V)
Terminators: V, mV, Cancel

Line: Settings are not available.

Trigger Level: To set the trigger level, press in the Trigger Level parameter field and
enter a value using the numeric keys. When a new value is entered, the terminator fly-
out will appear with the available units. Press on the desired units to terminate the entry.

Slope: To change the trigger slope, press on the Slope button to toggle between Rising
edge or Falling edge. Rising edge is the default setting.

Trigger Delay: To set the trigger delay, press in the Trigger Delay parameter field and
enter a value using the numeric keys. When a new value is entered, the terminator fly-
out will appear with the available units. Press on the desired units to terminate the entry.

SPA Mode:
Range: 0 ms to 65.5 ms (current sweep time to 65 ms for zero span)
Resolution: 0.1 ms (100 ns for zero span)
Default: 0 ms

FFT, FFT-WB Modes:
Range: 0 ms to 65 ms
Resolution: 100 ns
Default: 0 ms

VSA Mode:
Range: –65.5 ms to 65.5 ms
Resolution: 10 ns
Default: 0 ms