In table 5-8, Table 5-8 – Anritsu Series MS278XB User Manual

Page 239

Advertising
background image

Measurements

Operation Verification

MS278XB OM

5-83

26. Compute the Calculated Phase Noise for each value in Table 5-8.

27. Return Optimize Phase Noise to the Auto setting.

Table 5-8.

Single Sideband Phase Noise Test

Measured Carrier Power:

Center

Frequency

Offset

Measured

Measured Carrier to

Noise Ratio C/N (dBc)

Calculated Phase Noise

(dBc/Hz)

Specification

(dBc/Hz)

1.001 GHz

1 MHz

C/N–40 dB=

<

–139

1.005 GHz

5 MHz

C/N–40 dB=

<

–142

1.010 GHz

10 MHz

C/N–40 dB=

<

–142

Advertising