Front Panel Menus

Parameter Entry and Measurement Setup Dialog Archive

MS278XB OM

4-39

ACP Tab (2 of 2)

Alt2 Channel: Toggles the Alt2 channel
measurement off or on.

Alt2 Channel Width: Sets the bandwidth of the Alt2
channel.

Range: 1 Hz to 8 GHz
Default: 5 MHz

Alt2 Channel Spacing: Sets the spacing between
the main and Al2 channel.

Range: 1 Hz to 8 GHz
Default: 15 MHz

RRC Filter: This is used to turn the root-raised-cosine
filter (RRC) On or Off.

Rolloff Factor: Sets the roll-off factor for the
measurement (RRC filter must be on).

Range: 0.1 to 1.0 in steps of 0.01
Default: 0.22.

Bandwidth: Sets the bandwidth for the measurement
(RRC filter must be on).

Range: 1 Hz to 8 GHz
Default: 3.84 MHz.

Advanced Tab

/Hz: Toggles the result display between dBm or
dBm/Hz. The measurement type must be set to Abs
before /Hz can be toggled on.

Type: Toggles between an absolute or relative
measurement.

Noise Compensation: When noise compensation is
toggled on, a noise floor measurement is subtracted
from the channel power.

FFT: Toggles between FFT or Swept modes.

Start Time: Sets the start of the measurement
relative to a triggered event (currently not available).

Stop Time: Sets the stop time of the measurement
relative to a triggered event (currently not available).

Figure 4-15. Adjacent Channel Power Setup Dialog