10 

Pulse 

0 to 999999 

N/A 

Limit Switch 

If the switch has detective voltage, its range is +3V to +24V 

Note 1: When ambient temperature is (‐10 to 15℃and 25 to 50℃), temperature coefficients are: 
(1) Voltage, current measurement: ±(0.001%RD+0.0015%FS) /℃. 
(2) 2‐wire, 3‐wire, 4‐wire resistance measurement: ±(0.002%RD+0.001%FS)/℃. 
Note 2: Input features: 
(1) Voltage measurement: input impedance>1MΩ, overvoltage of ports is ±300VDC. 
(2) Current measurement: input impedance<10Ω, overvoltage of ports is ±1A DC. 

(3) Frequency、pulse  measurement:  Input  impedance>500MΩ,  square  wave,  low‐level  voltage  <0.3V,  high‐level 

voltage>2V, duty cycle is from 20% to 80%, overvoltage of ports is ±30VDC. 

(4) Resistance、RTD  measurement:  Measuring  current  is  about  0.7mA  at  400Ω  range,  and  about  0.3mA  at  4000Ω 

range, overvoltage of ports is ±30VDC. 

(5) Thermocouple、RTD measurement temperature: To be decided by the accuracy of mV measurement, resistance 

measurement, according to the 90 International temperature scale. 

(6) Range of cold junction automatically compensation: ‐10 to 50℃, accuracy: ±0.1℃, calibration is available. 