8.2.10 RTD Measurement (Temperature)

The  calibrator  supports11  common  RTDs  shown  in  Table  5‐4.  In 
addition,  calibrator  accepts  10custom  RTDs.to  add  a  custom  RTD 
please  refer  to  9.4.3.  The  calibrator  accepts  RTD  measurement 
inputs  in  two‐wire,  three‐wire,  or  four‐wire  connections  as  shown 
Figure  8.10.  The  four‐wireconnection  can  provide  the  highest 
measurement precision. 
  When  using  RTD  input  to  measure  temperature,  proceed  as 
follows: 
1. If  measurement  is  active,  press

once,  otherwise  press

twice. 

2. Press  the  navigation  keys  Up/Down  to  select  the  second  item 

from the pop‐up list, and then press Enteror Done to change the 
measurement  or  press  Options  to  view  the  setup  interface.  In 
setup  interface,  you  can  set  the  RTD  sensor  type,  connections 

and temperature units(℃, K,  ℉). 

3.   When RTD measuring, you can press Options to view the setup 

interface andset the measurement settings. 

4. To  get  higher  measurement  precision,  you  can  zero  measured 

value during resistance measurement. 

 

Figure 8.10 RTD Measurement 