(Figure  11.10  and  Figure  11.11).  Once  the  calibration  is  complete 
(Figure  11.11),  the  PV  range  adjustment  information  will  be 
displayed at the bottom of the interface. You can decide whether to 
adjust PV range. To reject, pressEsc or Reject;to adjust, press Enter 
orAdjust

11.4.4 Special Calibration

Special  Calibration  is  very  similar  to  the  PV  Calibration  operation.  The 
difference  is  that  Special  Calibration  adjusts  the  HART  transmitter’s  PV 
LRV and URV to get the coherence between the standard source ranges 
and loop current (CA, while digital analog output AOis described in the 
PV Calibration operation), so just refer to the PV Calibration operations 
for reference. 

11.4.5Output Trim

The  Output  Trim  adjusts  the  HART  transmitter’s  digital  analog  output 
according  to  the  actual  loop  current  measured  by  the  calibrator.  To 
perform an Output Trim, proceed as follows: 
1. In  the  HART  diagnostic  and  service  interface,  choose  Output  Trim 

from list, then you can pressEnter or Select to show the Output trim 
point selection interface. 

2. In this interface, you can select Low (zero point, 4mA) output point trim or High (span point, 20mA) output point 

 

Figure 11.11 HART PV Calibration result 