Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 19

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

12 

0.6mA).For  (1Ω  to  400Ω)  range,  1mA  driving  current  is  recommended,  and  for  (400Ω  to  4000Ω)  range,  0.1mA 
driving current is recommended. 

(5) Frequency,  pulse  source:  Source  square  wave  with  (50±10)%  duty  cycle,  amplitude  range  2V  to  12V,  amplitude 

accuracy 3%FS. 

(6) Loop 24V DC power: The maximum load current is 50mA, ripple <50mV. 
(7) Simulating  thermocouple,  RTD  source:  To  be  decided  by  the  accuracy  of  mV  source  and  resistance  source, 

according to the 90 international temperature scale. 

(8) All ports are short circuit protected. 
(9) The maximum overvoltage of all the sourcing ports is ±30VDC (less than 1 minute duration); the calibrator will be 

damaged when this range is exceeded. 

 

 