31 

junction  adjacent  to  the  thermocouple  jack.when  using  external  reference  mode,  The  fixed  reference  temperature 
value(range:  ‐10  to  50 ℃ )must  be  entered  manually  before 
thermocouple  measuring.  Figure  8.9  shows  the  thermocouple 
measurement connections.   
To  measure  temperature  by  using  thermocouple  simulation,  proceed 
as follows: 
1. If  measurement  is  active,  press

  once,  otherwise  press

 

twice. 

2. You can press the navigation keys Up/Downto select the first item 

from  the  pop‐up  list,  then  press  Enter  orDoneto  change  then 
measurement  or  pressthe  Optionsto  show  thermocouple 
measurement setup interface. In the thermocouple measurement 
setup  interface,  you  can  set  the  thermocouple  sensor  type, 

temperature units(℃, K,℉) and reference junction compensation 
methods. 

3. When thermocouple measuring, you can pressOptionsto show the 

thermocouple  measurement  setup  interface  to  set  the 
measurement settings. 

4. To  get  higher  temperature  measurement  precision,  you  can  zero 

measured value in millivolts measurement. 

 

Figure 8.9 Thermocouple Measurement 