Page 33

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

26 

2. If measurement is active, press

  once, otherwise press

twice. 

3. Press the navigation keys Up/Downto select the second item from the pop‐up list, and then pressEnter orDone to 

change the measurement. 

4. To improve measurement precision, you can short the volts input, then press Zero to clear the measured value. 

8.2.3 Frequency Measurement

Proceed as follows to measure frequency: 
1. Figure8.3 

shows 

the 

frequency 

measurement 

connections. 

2. If  measurement  is  active,  press

  once,  otherwise 

press

twice. 

3. Press  the  navigation  keys  Up/Downto  select  the  third 

item  from  the  pop‐up  list,  then  press  Enter  or  Done  to 
change the measurement. 

8.2.4 Pulse Counting

Proceed as follows to count pulse: 
1. Figure 8.4 shows the pulse counting connections. 
2. If  measurement  is  active,  press

  once,  otherwise 

press

twice. 

3. Press  the  navigation  keys  Up/Downto  select  the  fourth 

item  from  the  pop‐up  list,  then  press  Enter  orDone  to 

 

Figure 8.3 Frequency Measurement 