Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 35

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

28 

1. Figure 8.5 shows the resistance measurement connections. 
2. If measurement is active, press

  once, otherwise press

twice. 

3. Pressthe  navigation  keys  Up/Down  to  select  the  first  item  (range  0  to  400Ω)  or  second  item  (range  0  to  4000Ω) 

from the pop‐up list, then press Enter or Done to change the measurement or press Options to show resistance 
measurement configuration interface. 

4. To  improve  measurement  precision,  you  can  short  the  resistance  input,  then  press  Zero  to  clear  the  measured 

value. 

8.2.6 Limit Switch Test

Proceed as follows to test limit switch: 
1. Figure 8.6 shows the limit switch test connections. 
2. If measurement is active, press

  once, otherwise press

twice. 

3. Press  the  navigation  keys  Up/Down  to  select  the  third  item  from  the  pop‐up  list,  then  press  Enter  orDone  to 

change the measurement. 

4. Press View to view the switch test records with a maximum number of ten.A record displays the triggering state of   

the switch and the source value. 

8.2.7 Current Measurement

Proceed as follows to measure current: 
1. Figure 8.7 shows the current measurement connections. 

2. If measurement is active, press

once, otherwise press

twice. 

3. Press the navigation keys Up/Down to select the first item from the pop‐up list, then press Enter orDone to change 