8.2 Measuring

In the Basic Mode (as Figure8.1), if measurement is active, you can press switch keys (

,

,

,

,

)to pop 

up  relevant  measurement  items  list.  If  source  is  active,  you  can  press  a  switch  key  twice  to  pop  up  relevant 
measurement items lists. You can select a measurement from the list. To change any of measurement parameters, the 
measurement  must  be  in  active  state.  If  the  measurement  is  inactive,  press  the  navigation  keysUp/Downuntil 
measurement display area is focused. 

8.2.1 Millivolts Measurement

Proceed as follows to measure millivolts: 
1. Figure 8.2 shows the millivolts measurement connections. 
2. If  measurement  is  active,  press

once,  otherwise  press

twice. 

3. Press  the  navigation  keys  Up/Downto  select  the  first  item 

from  the  pop‐up  list,  and  then  pressEnter  or  theDone  to 
change the measurement. 

4. To  improve  measurement  precision,  you  can  shortthe 

millivolts input, and then press Zero to clear the measured 
value. 

8.2.2 Volts Measurement

Proceed as follows to measure volts: 
1. Figure 8.2 shows the volts measurement connections. 

Figure 8.2 Millivolts or Volts Measurement 