11. HART Communication (only HART is available) 

11.1 Brief and note

Thecalibrator is a secondary master device, and has no impact on the normal communication between a primary 
master device and a HART device. The jacks of HART (refer to Table7‐1) have a built‐in HART resistor. This HART resistor 
is optional and its rate current is 44mA. 
Note 1: The calibrator can measure current only with a range of ‐30 to 30mA. If loop current (CA) is unnecessary, you 
can bypass the current measure jacks, also , the calibrator can poll over 15 devices (address 1 to 15) with an external 
HART resistor. 
Note 2: The HART jacks can endure ±50V when an external HART resistor is being used. 
In the calibrator HART mode, the calibrator will supply these operations as follows: 
 SupportsHART  process  variables  updating.  The  calibrator  can  show  primary  variable  (PV),  rang  (%),  PV  current 

value  (AO),  loop  current  (CA),  thesecondary  variable,  the  third  variable,  and  the  fourth  variable.  Thesecondary 
variable, the third variable, and the fourth variable are special process variables definedby a transmitter. 

 Support  HART  diagnostic  and  service:  HART  diagnostic  and  service  operations  will  allow  you  to  calibratea 

supported HART instrument. 

 Support  HART  setup:  HART  setupoperations  allow  you  to  configure  common  parameters  of  a  supported  HART 

instrument. 

 

 