Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 81

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

74 

11.4.3 PV Calibration

Adjust  the  HART  transmitter’s  PV  LRV  and  URV  to  get  the  coherence 
between  the  standard  source  range  and  digital  analog  output  (AO).  To 
perform the PV Calibration operation, proceed as follows: 
1. In  the  HART diagnostic  and  service interface, choose  PV Calibration 

from list, then pressEnteror Select to show the PV Calibration Setting 
Interface (Figure 11.9). 

2. In the SettingInterface, you can set the sensor type and the range of 

the transmitter. If the device is a pressure transmitter, the pressure 
module has to be connected to the calibrator and pressure controller 
such  as  hand  pump  is  also  available  for  this  application.  Note:  the 
UUT’s input pressure unitshould match the previousUnit. The unit of 
pressure will be automatically selected by the calibrator. 

3. Press  Enter  or  Done  to  view  the  PV  CalibrationProcess  interface 

(Figure  11.10).  Note  that  the  calibrator  will  judge  whether  these 
input values are available for the transmitter before starting the PV 
Calibration Process. 

4. The  PV  calibration  process  allows  you  to  calibrate  the  Lower  (0%) 

and Upper (100%) Range Values (LRV and URV) have been previously 
set in PV Calibration Setting interface. Located at the top of the PV Calibration interface is the error summary table 

 

Figure 11.10 HART PV Calibration Process 