7.2 Introduction of Jacks Function

Table 7‐1 

Jacks

Introduction

2& 9 

①Voltage  measurement②current  measurement  ③resistance  measurement  (2W)④
limit switchtest  ⑤pulse measurement  ⑥frequency measurement 

2, 9 & 10 

Resistance measure (3W) 

1, 2, 9 & 10 

Resistance measure (4W) 

1 & 10 

HART communication (

only HART is available

4 & 7 

①Voltage source②current source (external power supply)③pulse source 
④frequency source   

4 & 5 

Current source (internal 24V dc power supply) 

3 & 8 

Resistance source 

5 & 6 

24V dcloop source 

11 

Jack for thermocouplemeasurement 
The calibrator can provide automaticcold junction compensation function when this jack 
is used 

12 

Jack for simulating thermocouple 
The calibrator can provide the automatic cold junction compensation function when this 
jack is used 

Pressure module 

Connecting CDP pressure modules 