9.4.2 Simulate Transmitter

The  transmitter  is  a  device  which  can  measure  temperature,  electrical  signals  and  pressure,  etc.  Generating  a  signal 
according to the measurement imposed. 
The  calibrator  can  simulate  a  transmitter.It  can  measure  RTDs,  thermocouples,  frequency,  volts,  mA,  ohms  and 
pressure (when the pressure module is connected), and output mA and voltssignal. 
Before starting the transmitter simulation, you should ensure the external device is available. 
Taking the example of an RTD Temperature Transmitter (T/I), proceed as follows: 
1. Select the input and output quantities of the transmitter. 
2. Enter the input and output spans of the transmitter. 
3. Start running the simulate transmitter. 
On the Quantity page, select mA&24V from the output quantity list and select RTD from input quantity list. In addition, 
choose the correct sensortype, temperature unit and connection wires for the RTD (as shown in Figure 9.2). 
On the Range page, set the input range 0 to 200℃  and the output range 4 to 20mA. Sometimes, it is necessary to set 
the transfer function: Linear/Square root (as shown in Figure 9.3).   
Press  Enter  or  Start  to  start  running  the  transmitter  simulation.  The  simulation  will  pause  when  the  transmitter  is 
working. When the simulation is paused, both the input and output of the transmitter will also pause   