9.4.3 Custom RTDs Library

The calibrator can measure temperature with most RTDs. But sometimes there is a need to support non‐standard RTDs 
and Standard PRTs (standard platinum resistance thermometers). 
The calibrator allows you to input the customized values, so you can measure temperature usingevery kind of RTDs and 
Standard PRTs. 
How to create a custom industry RTD?
To  get  high  precision,  IEC  and  ASTM  standards  specify  a  more  complex  polynomial  that  fine‐tunes  the 
resistance‐temperature relationship. One form of that equation is given here as below. The industry RTDs also specify 

values for the constants R0, A, B, C (The C constant is only used for temperature less than 0℃). 

1

100

 

For example, for the common Pt385 RTD with 100Ω resistance at 0 °C, the values are: 

100Ω 

A

3.9083

10  

B

5.775

10  

C

4.183

10

 

How to create a standard PRT?
The  Standard  Platinum  Resistance  Thermometer  (SPRT)  Calibration  Laboratory  of  the  NIST  ThermometryGroup  can 
distinguish  among  the  ITS‐90  from  the  argon  triple  point  (Ar  TP,  –189.3442  °C)  to  the  silver  freezing  point  (Ag  FP, 
961.78 °C) for the calibration of SPRTs. 
The  ITS‐90  defines  temperature  through a  set  of specified  thermometric  fixed points,  interpolation  instruments,  and 