76 

trim, then you  can press  Enter  or  Done to display  the Output  Trim 
interface (as shown in Figure 11.12). 

3. In  the  Output  Trim  interface,  press  Enteror  Trim  to  execute  the 

output point adjustment. The trim value can be fetched by pressing 
Fetch, or it can be inputted manually. 

4. If Output Trim is successfully executed, the calibrator will goes back 

to the Output Trim point selection interface. You can repeat steps 2 
and 3 to start another Output Trim. 

 

 

 

Figure 11.12 HART Output Trim 