11.5 HART Setup

Whenthe  measurement  is  active  in  HART  communication  mode,  pressSetupto  show  the  HART  setup  interface.The 
calibrator provides five setup functions (refer to table 11‐1). 

Table 11‐1 

Setup

Variables

Attributes

Description

Basic

Device Label 

Revisable 

English  input  is  available,  no  more  than  8 
characters 

Primary variable range units 

Revisable 

Can  be  changed  into  any  units  supported  by 
transmitter 

Primary variable range lower 
limit 

Revisable 

The  setting  range  cannot  be  less  than 
transmitter range lower limit 

Primary 

variable 

range 

upper limit 

Revisable 

The setting range cannot exceed the transmitter 
range upper limit 

Damping time 

Revisable 

Float type,the value cannot be less than 0 

Transfer function 

Revisable 

Can  be  changed  into  linear  transfer  or  rooting 
transfer 

Sensor

SensorS/N 

Unchangeable 

N/A 

Sensor range lower limit 

Unchangeable 

N/A 

Sensor range upper limit 

Unchangeable 

N/A 

Sensor minimum span 

Unchangeable 

N/A 