16 

Simulating RTD Source and using RTD to Measure Temperature Accuracy. (Working temperature: 20

±5)

Table5‐4

RTD Measurement and Source accuracy

Measure and

Simulate

Standard

Temperature Range ()

Accuracy ()

Measure

2W3W

Measure

4W

Source

Pt10(385) 

IEC 751 

‐200 to 850 

‐200 to 200 

0.65 

0.60 

0.65 

200 to 600 

0.82 

0.72 

0.82 

600 to 850 

0.96 

0.82 

0.96 

PT100(385) 

IEC 751 

‐200 to 850 

‐200 to 200 

0.15 

0.1 

0.15 

200 to 600 

0.26 

0.16 

0.26 

600 to 850 

0.34 

0.20 

0.34 

Pt100(3916) 

IEC 751 

‐200 to 850 

‐200 to 200 

0.15 

0.1 

0.15 

200 to 600 

0.26 

0.16 

0.26 

600 to 850 

0.33 

0.20 

0.33 

Pt200(385) 

IEC 751 

‐200 to 850 

‐200 to 200 

0.37 

0.32 

0.69 

200 to 600 

0.51 

0.41 

0.92 

600 to 850 

0.61 

0.48 

1.08 