Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 77

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

70 

aborted immediately and show the HART device list. Otherwise calibrator will continue searching from address 1 to 
15, and automatically stop polling at address 15. In the period of polling, you can press Stop tostop polling. If more 
than  one  hart  device  is  found,  calibrator  will  show  the  HART  devices  list,  otherwise  calibrator  will  prompt  you 
whether to poll HART device again. A maximum of 15 HART devices 
can be supported. 

3. Press  Enter  or  Select  to  Select  a  HART  device  from  the  list,  the 

calibrator  will  connect  to  the  selected  HART  device,  and  the  HART 
device information will be displayed (As shown in Figure 11.3). 

4. PressEnter  or  Done  ,  then  the  calibrator  will  go  back  to  the  main 

screen in Basic Mode, the HART communication interface displayed 
in the measurement display area, and the HART icon will be shown 
in the status bar (Figure 11.4). To press Esc key or the Back softkey, 
you can return to the HART device list to select another HART device 
to get the device information. 

Whenthe  HART Communication Interface is  active,  the  Process, Service
and Setup softkeyare  available.  You  can  press  the  softkey  to  access  to 
the corresponding HART operations. 

11.3 HART Process Variables Update

Whenthe  HART  Communication  Interface  is  active,  you  can 
pressProcesstoshow  HART  Process  Variables  interface  (Figure  11.5).  In 

 

Figure 11.5 HART Process Variables 