42 

switch source to current.   

4. When sourcing current with internal power, the 24VDC loop power states must remain as ON. If the 24VDC power 

state is OFF, it will be automatically turned back ON states after switching source to current with internal power. 

5. Refer to the  sections 8.3.1,  8.3.2,  8.3.3,  8.3.4  or  8.3.5,  select  a  method  to  change  the  source value  according to 

your need. 

8.3.12 Pressure source (only available for pressure

module)
Proceed as follows to pressure source: 
1. Figure 8.16 shows the pressure source connections. 
2. Connect the pressure module. 

3. If  source  is  active,  press

once,  otherwise  press

twice. 

4. Press  the  navigation  keys  Up/Down  to  select  the  first 

item  from  the  pop‐up  list,  then  press  Enter  orDone  to 
change the source to pressure. 

5. You can press Units to view the pressure units list which 

are  supported  by  the  calibrator.  You  can  change  the 
pressure unit. 

6. You  can  press  Module to  view  the  information  about 

the pressure module. 

 

Figure 8.16 Pressure Source 