When a pressure module is connected, the pressure module 
icon will appear in the status bar and beep at the same time. 
Figure 8.8 shows thepressure measurement connections. 
Proceed as follows to measure pressure: 

1. If  measurement  is  active,  press

once,  otherwise 

press

twice. 

2. Press  the  navigation  keys  Up/Down  to  select  the  first 

item  from  the  pop‐up  list,  then  press  Enter  or  Done  to 
change the measurement. 

3. You can press Unitto pull up the pressure unit list which 

issupported by the calibrator. 

4. You  can  press  Module  to  view  the  pressure  module’s 

information 

5. To  improve  measurement  precision,  you can  press  Zero 

to clear the measured value. 

8.2.9 Thermocouple Measurement (Temperature)

The calibrator supports thirteen standard thermocouples; thermocouple types are identified bycharacter: S, R, B, K, N, 
E, J, T, C, D, G, Land U. Table5‐3 shows the temperature ranges and accuracies of these thermocouples. Thermocouple 
measurement accepts two reference junction compensation methods: internal reference mode or external reference 
mode.  When  using  the  internal  reference  mode,  the  calibrator  will  automatically  measure  the  temperatureof  cold 

 

Figure 8.8 Pressure Measurement 