Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 44

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

37 

Figure 8.13 Ramping Configuration 

main screen. 

4. Press Auto to beginautomatic stepping, and the softkey label will change to Pause
5. Press Pause to pause automatic stepping. The softkey label will change to Resumeat the same time.You can press 

Resume to continue automatic stepping. 

If you want to exit automatic stepping, press Esc orExit

8.3.5 Ramping the source value

In  the  Basic  Mode,  the  calibrator  offersthe  source  value  ramping. 
Ramping can smoothly and continuously increase or decrease the source 
value.  The  size  of  steps  is  determined  by  the  start  value,  end  value  and 
the ramp time. Before ramping, you need to configure the ramp settings. 
Each  source  item  has  its  own  ramp  settings.  When  source  is  active  and 
source  item  is  not  pulse,  frequency  and  pressure,  you  can  press  Ramp 
toview  the  ramp  setup  interface.  Figure  8.13  and  Figure  8.14  are 
examples of ramp settings and the ramp results of settings. 
The  Start  value  cannot  equal  the  End  value,  and  their  values  cannot 
exceed the souring item’s output range.   
Note: The ramp time is a single rising time, also is a single falling time, so 
the ramp rising process and the falling process is symmetric. In the case 
of  fixedstart  value  and  end  value,  the  longer  ramp  time  it  takes  the 
slower  ramp  and  smaller  steps size  it will  be.  The  range of  ramp time is 