this interface you can view more HART Process Variablesincluding primary variable (PV), rang (%), PV current value (AO), 
loop current (CA), thesecondary variable, the third variable, and the fourth variable, and their values are continuously 
updated.  PV,  %,AO,  and  CA  are  selectable;  you  canchoose  a  process 
variable and return to the Communication Interface (Figure 11.4). 

11.4HART Diagnostics and Service

Whenthe measurement is active in HART Communication Interface, 

you can pressService todisplay the HART diagnostic and service interface 
(Figure  11.6).  The  calibrator  provides  Loop  Test,  PV  (Primary  Variable) 
Zero Trim, PV Calibration, Output Trim and Special Calibration operations 
to  calibrate  a  transmitter.Note  that  Loop  Test  and  Output  Trim  are  not 
available if the transmitter is in multi‐drop mode. 

11.4.1 Loop Test

The Loop Test operation will send a command to the transmitter to 

set  its  analog  output  to  the  specified  value.  To  perform  a  Loop  Test 
operation, proceed as follows: 
1. In the HART diagnostic and service interface, choose Loop Test from 

list,  and  then  you  can  press  Enter  or  Select to  show  Loop  Test 
interface (Figure 11.7); 

2. In this interface, you can use numeric keys to input valueof the test 

current.  Its  range  is  4  to  20mA.  Then  pressEnter  or  Done  to  start 

 

Figure 11.6 HART Diagnostics& Service 