Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 69

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

62 

 

Figure 10.7 AsCal Calibration Method 

The Calibration Method page will appear when the task can be calibrated 
both manually and automatically (as shown in Figure 10.7). 
About Automatic Calibration
Automatic calibration is possible when the calibrator is able to 
generate/simulate the UUT’s input signal, and measure the UUT’s output 
signal. 
The Step Time (set point delay) setting is in use only when the 
Calibration Method is set to Automatic. It determines theinterval 
between the two calibration points. 
If the UUT is a limit switch, this page will display the ramping time. 
About Manual Calibration
In Manual Calibration, you have to manually set the value that the 
calibration point requires. Manual calibration is possible in almost all 
situations, even for calibrations that can be done automatically.  
 

 