Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 57

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

50 

interpolation equations. 
The calibrator uses the first and last deviation function for the calculation for SPRT.If you use other temperature ranges, 

  and 

can be assigned to zero (i = 7, 8, 9, 10, 11). 

The table 9‐1 list the sub ranges, required calibrated in the NIST SPRT Laboratory. 

Table 9‐1 

Temperature range(

)

Required Fixed Points

Deviation Function

-189.3442 to 0.01

Ar TP, Hg TP, TPW

0 to 29.7646 

TPW, Ga TP 

1  

0 to 156.5985 

TPW, GaTP,In FP 

1  

0 to 231.928 

TPW, In FP, Sn FP 

1

9

1

2

 

0 to 419.527 

TPW, Sn FP, Zn FP 

1

8

1

2

 

0 to 660.323

TPW, Sn FP, Zn FP, Al FP

On the RTDs Management page, you can delete an existing RTD by pressing Deleteand re‐edit it by pressing Enter or 
Edit. 
When the custom RTD is added to the RTDs Library successfully, your calibrator is now ready to make measurements 
using your custom RTD. 