Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 18

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

11 

Electrical Signal SourceSpecifications: (working environment: 20±5, accuracy for one year)

Table5‐2

Function

Range

Resolution

Accuracy

Voltage DC 

‐10.000 to 75.000 mV 

1μV 

0.02%RD+0.005%FS 

0 to 12.0000 V 

0.1mV 

0.02%RD+0.005%FS 

Current DC 

0 to 22.000 mA 

1μA 

0.02%RD+0.005%FS 

Resistance 

1 to 400.00 Ω 

10mΩ 

0.02%RD+0.005%FS 

1 to 4000.0 Ω 

100mΩ 

0.03%RD+0.01%FS 

Frequency 

0 to 50000.0 Hz 

0.1Hz 

0.005%RD+0.002%FS 

Pulse 

0 to 999999 

N/A 

DC24V 

N/A 

N/A 

0.5V 

Note 1: When ambient temperature is (‐10 to 15℃ and 25to 50℃), temperature coefficients are: 

(1) Voltage, current, resistance source: ±(0.001%RD+0.0015%FS) /℃. 
Note 2: Source features: 
(1) Millivoltssource: Max load capability is 1mA, load variance ratio less than 50μV/mA. 
(2) 12V source: Max load capability is 5mA, load variance ratio less than 1mV/mA. 
(3) Current  source:  Internal  24V  power  supply,  max  load  capability  is  20mA@1kΩ;  External  power  supply,  the  max 

voltage is 36V. 

(4) Simulating  resistance:  (1Ω  to  110Ω)exciting  current  (0.6mA  to  4mA),  (110Ωto  400Ω)exciting  current  (0.3mA  to 

4mA),  (400Ω  to  1000Ω)exciting  current  (0.06mA  to  1.5mA),  (1000Ωto  4000Ω)exciting  current  (0.01mA  to 