Configure gpib interface – GW Instek PSW-Series Programming & Command manual User Manual

Page 31

Advertising
background image

REMOTE CONTROL

31

Configure GPIB Interface

To use GPIB, the optional GPIB to USB (GUG-001) adapter must be
used. The GPIB to USB adapter must be connected before the PSW
is turned on. Only one GPIB address can be used at a time.

Configure GPIB

1. Ensure the PSW is off before proceeding.

2. Connect the USB cable from the rear panel

USB B port on the PSW to the USB A port on
the GPIB to USB adapter.

3. Connect a GPIB cable from a GPIB controller to

the GPIB port on the adapter.

Type B plug for

PSW series

Type A plug

From

computer

GUG-001

4. Turn the PSW on.

5. Press the Function key to enter the

Normal configuration settings.

Page 21


Set the following GPIB settings:

F-22 = 1

Set the rear panel USB port to
GPIB-USB (GUG-001)

F-23 = 0~30

Set the GPIB address (0~30)

GPIB constraints

Maximum 15 devices altogether, 20m cable
length, 2m between each device

Unique address assigned to each device

At least 2/3 of the devices turned On

No loop or parallel connection