REMOTE CONTROL

75

Trigger Commands

The trigger commands generate and configure software triggers.

TRIGger:TRANsient[:IMMediate] ...................................... 75

TRIGger:TRANsient:SOURce ............................................ 75

TRIGger:OUTPut[:IMMediate] .......................................... 76

TRIGger:OUTPut:SOURce ................................................. 76

Trigger Command Examples ............................................... 76

TRIGger:TRANsient[:IMMediate]

Set

Query

Description

Generates a software trigger for the transient
trigger system. On a trigger, sets the voltage &
current. Refer to the :CURR:TRIG and VOLT:TRIG
commands on page 69 and 73, respectively.

Syntax

TRIGger:TRANsient[:IMMediate]

Related
Commands

[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]

TRIGger:TRANsient:SOURce

Set

Query

Description

Sets or queries the trigger source for the transient
system.

Syntax
Query Syntax

TRIGger:TRANsient:SOURce {BUS | IMMediate}
TRIGger:TRANsient:SOURce?

Parameter/Return BUS

Internal software trigger. Waits for the
*TRG (or IEEE 488.1 “get” group
execute trigger) command to start the
trigger.

IMMediate

Starts the trigger immediately.
(default)

Example

TRIG:TRAN:SOUR BUS
Sets the trigger source as BUS.