CC (Constant current mode)

Indicates if the PSW is in CC
mode.

10

1024

OND (Output ON Delay)

Indicates if Output ON delay time
is active

11

2048

OFD (Output OFF Delay)

Indicates if Output OFF delay
time is active

12

4096

PR (Program Running)

Indicates if a Test is running

13

8192

Condition
Register

The Operation Status Condition Register
indicates the operating status of the power
supply. If a bit is set in the Condition register, it
indicates that the event is true. Reading the
condition register does not change the state of
the condition register.

PTR/NTR Filters

The PTR/NTR (Positive/Negative transition)
register determines the type of transition
conditions that will set the corresponding bit in
the Event Registers. Use the Positive transition
filter to view events that change from false to
positive, and use the negative transition filter to
view events that change from positive to
negative.

Positive Transition

0→1

Negative Transition

1→0

Event Register

The PTR/NTR Register will dictate the type of
transition conditions will set the corresponding
bits in the Event Register. If the Event Register
is read, it will be cleared to 0.