System:error, System:keylock:mode, System:klock – GW Instek PSW-Series Programming & Command manual User Manual

Page 88

Advertising
background image

PSW Series Programming Manual

88

Parameter/
Return parameter

0

Disable

1

GPIB-USB adapter

2

Auto detect speed

3

Full speed only

Example

SYST:COMM:USB:REAR:MODE 1
Sets the rear panel USB-B port mode to GPIB-USB
adapter.

SYSTem:ERRor

Query

Description

Queries the error queue. The last error message is
returned. A maximum of 32 errors are stored in the
error queue.

Query Syntax

SYSTem:ERRor?

Paramter/Return <NR1>,<string>

Returns an error code followed by
an error message as a string. The
string is returned as “string”.

Example

SYSTem:ERRor?
-100, “Command error”

SYSTem:KEYLock:MODE

Set

Query

Description

Sets or queries the key lock mode. This setting is
the equivalent of the F-19 function setting.

Syntax
Query Syntax

SYSTem:KEYLock:MODE {0|1}
SYSTem:KEYLock:MODE?

Parameter /
Return parameter

0

Panel lock: allow output off.

1

Panel lock: allow output on/off.

SYSTem:KLOCk

Set

Query

Description

Enables or disables the front panel key lock.

Syntax
Query Syntax

SYSTem:KLOCk { OFF | ON | 0 | 1}
SYSTem:KLOCk?

Parameter

0

Panel keys unlocked

Advertising