REMOTE CONTROL

71

Example

SOUR:CURR:SLEW:RIS 72
Sets the rising current slew rate to 72A/s.

[SOURce:]CURRent:SLEW:FALLing

Set

Query

Description

Sets or queries the falling current slew rate. This is
only applicable for CC slew rate priority mode.

Syntax

[SOURce:]CURRent:SLEW:FALLing {<NRf>|MIN|MAX}

Query Syntax

[SOURce:]CURRent:SLEW:FALLing? [MIN|MAX]

Parameter/Return NRf

0.01A/s~72.00A/s (PSW 30-36)
0.1A/s~144.0A/s (PSW 30-72)
0.1A/s~216.0A/s (PSW 30-108)
0.01A/s~27.00A/s (PSW 80-13.5)
0.01A/s~54.00A/s (PSW 80-27)
0.01A/s~81.00A/s (PSW 80-40.5)
0.01A/s~14.40A/s (PSW 160-7.2)
0.01A/s~28.80A/s (PSW 160-14.4)
0.01A/s~43.20A/s (PSW 160-21.6)
0.001A/s ~ 9.000A/s (PSW 250-4.5)
0.01A/s ~ 18.00A/s (PSW 250-9)
0.01A/s ~ 27.00A/s (PSW 250-13.5)
0.001A/s ~ 2.880A/s (PSW 800-1.44)
0.001A/s ~ 5.760A/s (PSW 800-2.88)
0.001A/s ~ 8.640A/s (PSW 800-4.32)

MIN

Minimum falling current slew rate

MAX

Maximum falling current slew rate

Example

SOUR:CURR:SLEW:FALL 1
Sets the falling current slew rate to 1A/s.

[SOURce:]RESistance[:LEVel][:IMMediate]
[:AMPLitude]

Set

Query

Description

Sets or queries the internal resistance in ohms.

Syntax

[SOURce:]RESistance[:LEVel][:IMMediate][:AMPLitude]
{<NRf>|MIN|DEF|MAX|?}