Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 13

Advertising
background image

11

The operator can program several parameters in the calibrator

which are further detailed in this manual. Programmable parame-

ters include:

• Date and time
• Owner name
• Auto off time
• Port selection
• Select Voltage (v) or current (mA) monitoring
• Enable/Disable 24VDC power supply
• Zero modules
• Tare
• Damping
• Percent error function
• Engineering units select
• User defined unit of measure
• Dual sensor mode
• Pressure or Temperature switch test mode
• Calibration of base unit
• Module calibration
• Datalogging via USB
• Datalogging to internal memory
• Transfer datalog to SD card

Owner name and base unit calibration functions are password

protected. The passwords may be obtained by contacting the fac-

tory or downloading the “HHC Secure” application from the web

site. “HHC Secure” will generate the HHC’s unique pass codes

based on user input of the base unit serial number. Make a note

of these pass codes

Owner Pass code_______________________________

Calibration pass code _________________________________

Quick-select modules are available for measuring a wide variety

of pressure and temperature ranges. They can be “hot swapped”

within the base unit while it is powered on without causing any

damage to the base unit or the module. “Hot Swap” refers to

installing or removing modules while the base unit is powered

on. The base unit will recognize the module and display the unit

of measure in which the module was originally calibrated and

labeled. Quick-Select pressure modules are available in accura-

cies from ±0.1% to ±0.025% of span.

SECTION 5

HOW TO OBTAIN PROGRAMMING

PASSCODES

SECTION 6

QUICK-SELECT MODULE

OVERVIEW

SECTION 4

BASE UNIT OVERVIEW CONT.

USB PORT

SD CARD PORT

USB MICRO B
CONNECTOR