SECTION 15 g.

SWITCH TESTING FUNCTION

CONT.

Use DOWN or UP ARROW to highlight “Measurement Channel…”

Press ENTER

Use DOWN or UP ARROW to highlight “CH1” (Channel 1

module) or “CH2” (Channel 2 module)…

Press ENTER

HHC will go back to Switch Testing MENU

Use DOWN or UP ARROW to highlight “Signal Type…”

Press ENTER

Use DOWN or UP ARROW to highlight “Current or Voltage…”

Press ENTER

Use DOWN or UP ARROW to highlight “Switch Type…”

Press ENTER

Use DOWN or UP ARROW to highlight “Normally Open or Nor-

mally Closed…”

Press ENTER

Display will briefly display the message “Begin Test” then immedi-

ately enter switch test mode.

In Switch Test Mode the display will initially show the Live pres-

sure (pressure being applied at that time), and units of measure.

The switch trip point, reset point and dead band (DB) will be

blank indicated by “---“
Setup is now complete and switch testing can begin. Depending

on the options chosen above either apply additional pressure or

release pressure slowly.
Note: depending on number of modules installed the update rate

will be either 0.2 or 0.3 seconds. A rapid increase or decrease

in pressure near the switch setpoint can result in inaccurate

readings
When the switch opens or closes the pressure/temperature read-

ing and units of measure will be displayed as “trip” point.
Slowly release or increase pressure (opposite of the action taken

in the earlier step) until switch state changes again. This value

will be displayed as the “reset” value along with units of measure.

The deadband (DB) will also be calculated and displayed. This

value is the difference between the trip point and reset point.

If desired, record the readings displayed on the screen. These

values are not included in the manual Datalog.
To reset switch test and maintain the same parameters press the

ZERO key to return to initial testing screen.
To exit switch test mode press ESC
To enter Main MENU Programming Mode press MENU.