Enter ch1 tare – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 23

Advertising
background image

21

Exit Tare Mode
Press TARE key
Press PORT SELECT key repeatedly to select Channel 1,

Channel 2, or both active pressure modules
Press ENTER
Selected modules are now in Measurement Mode without a tare

value.
Program Manual Tare Value:
Press MENU key
Press DOWN or UP ARROW key to highlight “Modes…”
Press ENTER
Press DOWN or UP ARROW key to highlight “Tare…”
Press ENTER
Press DOWN or UP ARROW key to highlight “CH1 auto/manual

or CH2 auto/manual”
Press ENTER
Press DOWN or UP ARROW key to highlight Auto or Manual
Select “manual” and press “ENTER” to advance to “Enter CHx tare”

Press left or RIGHT ARROW key to select digit
Press UP and DOWN ARROW keys to advance numeric values
When desired tare value is displayed
Press ENTER
The HHC will enter Measurement Mode.
Select Automatic Tare Mode
Note that the default tare mode is Automatic. Tare Mode will be

reset to automatic when the unit is turned off. It is unnecessary to

program automatic tare mode unless manual mode was used in

the same session.
Press MENU key
Press DOWN or UP ARROW key to highlight “Modes…”
Press ENTER
Press DOWN or UP ARROW key to highlight “Tare…”
Press ENTER
Press DOWN or UP ARROW key to highlight “CH1 auto/manual

or CH2 auto/manual”
Press ENTER
Press DOWN or UP ARROW key to highlight Auto or Manual
Select “auto” then press ENTER to exit back to Measurement

Mode

SECTION 14 h.

TARE FUNCTION CONT.

-0.24

0.000

–0.000

PSi

Tare

inH20

mA

Enter CH1 tare

00020.00000

^

psi

Enter to accept

Esc to cancel