Tc manual ref, Tc module, Tc junction type – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 48

Advertising
background image

46

Use DOWN or UP ARROW to highlight “Channel…”

Press ENTER

Use DOWN or UP ARROW to highlight “CH1” “CH2”

Press ENTER
Note: If the incorrect type of module is installed a “MODULE

ERROR” message will be shown.
Return to TC Module screen

Use DOWN or UP ARROW to highlight “Junction type”…

Press ENTER

TC Junction type screen appears.

Use DOWN or UP ARROW to scroll through Thermocouple types

J,K, T, E, R, S, B,N “

When the desired TC type is displayed

Press ENTER

Note; By selecting the thermocouple type the PTE-2 will access

the factory programmed coefficients in the module which will

allow the PTE-2 to display in a unit of temperature

Return to TC Module screen
Use DOWN or UP ARROW to highlight “Auto/manual ref…”
Press ENTER
Use DOWN or UP ARROW to highlight “Auto or Manual…”
By selecting the “Auto” mode, the calibrator will utilize the internal

thermistor/resistor based on the reference junction that is resi-

dent inside the interface module.
The “Manual” mode allows for the use of an external reference

junction. A stable temperature reference, usually in the form of

an ice bath, is required.
Press ENTER
If the “auto” mode was selected, the calibrator will now return to

TC Module screen with “Resolution” highlighted, skip to Resolu-

tion programming.
If Manual was selected continue below

TC Manual Ref

00000.00

^

Enter to accept

Esc to cancel

TC Module

Channel…

Junction type…

Auto/manual ref…

Resolution…

Apply settings

TC Junction Type

Thermocouple

Type:

J

Fe-Cu/Ni

SECTION 15 k.

THERMOCOUPLE TEMPERATURE

MEASUREMENT CONT.

Ꮨƽ