Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 24

Advertising
background image

22

SECTION 14 j.

WIRING DIAGRAMS

+

+

24Vdc Out

0-33Vdc In

0-22mA In

COM

24Vdc Out

0-33Vdc In

0-22mA In

COM

24Vdc Out

0-33Vdc In

0-22mA In

COM

RESISTOR*

EXTERNAL

24 VOLT DC

SWITCH HOOKUP

USING INTERNAL

SUPPLY SWITCH

NORMALLY OPEN

SWITCH HOOKUP

USING INTERNAL

SUPPLY SWITCH

NORMALLY OPEN

EXTERNAL

24 VDC SUPPLY

Note that you can use a normally closed switch as well

*Resistor should be sized to allow current as follows:

2 mA < I < 36 mA, 2 mA Ȃ 12 k , 35 mA Ȃ 686

SWITCH HOOKUP

USING INTERNAL

SUPPLY SWITCH

NORMALLY OPEN

4/20 TRANSMITTER HOOK UP USING,

INTERNAL SUPPLY

4/20 TRANSMITTER HOOK UP USING,

EXTERNAL SUPPLY

4/20

TRANSMITTER

24 Vdc

EXTERNAL

SUPPLY

+ –

4/20ma

TRANSMITTER

+ –

+

+24V ma Comm Vdc

+24V ma Comm Vdc

V OUT TRANSDUCER HOOK UP USING,

INTERNAL SUPPLY

V OUT TRANSDUCER HOOK UP USING,

EXTERNAL SUPPLY

V OUT

TRANSDUCER

24 Vdc

EXTERNAL

SUPPLY

+ –

V OUT

TRANSDUCER

+

+24V ma Comm Vdc

+24 ma Comm Vdc

̈

̈

+

SECTION 14 i.

mA/V FUNCTION

The HHC has the capability to monitor either DC milliamp current

0-22 mA or DC voltage 0-33 Vdc. Use of this feature requires

connection to the voltage or current source via the 4mm “Banana

Jacks” located above the units LCD. Note that this calibrator can

monitor only one parameter, either voltage or current, at a time.
To monitor voltage connect the negative lead of the voltage

source to the socket labeled “COM” with a black ring surrounding

the socket. Connect the positive lead of the voltage source to the

red ringed terminal labeled “0-33Vdc In”.
To monitor current connect the negative lead to the black ringed

terminal labeled “COM. Connect the positive current source to

the red ringed terminal labeled “0-22mA In”.
Devices under test may be powered by an external power source,

or by the 24 Vdc power supply built into standard (non-intrinsically

safe) HHC base units. See wiring diagrams in section 14j.
Press the mA/V key to activate voltage and current monitoring.

Press again to rotate through Volts, mA and pressure only

measurement.