SECTION 14 h.

TARE FUNCTION

To close a Datalog and prepare the HHC to begin a new Datalog

press the ESC key. The next press of the DATA STORE key will

begin a new Datalog and will continue to record data in the new

log with every subsequent press of the key.
Up to 64 individual logs can be stored in the on board memory

with a total of 16,000 individual records.

The HHC has the ability to Tare (subtract an operator selected

value) from the displayed pressure measurement value of either

one or two installed Quick-Select modules. This function is appli-

cable to pressure measurements only. This is most commonly

required in applications where a pressure pre-load, such as in a

weighing application, must be subtracted from the displayed pres-

sure measurement. If the HHC is being used with two Quick-Se-

lect modules, independent tare values can be entered for each

of the pressure modules. When tare function is active bar graph

displays will indicate actual percentage of full scale pressure not

the tare value. When the HHC is in tare mode the word “tare” will

appear below the engineering units.
There are two methods to select a tare value, automatic and

manual. Automatic Tare Mode will subtract the actual pressure

applied to the module at the time tare function is activated.

Manual Tare Mode allows the operator to enter a pressure value

to be subtracted from the actual pressure applied to the module.

Automatic tare is the default condition. If Manual Tare Mode is

desired go to the section named Program Manual Tare Value

below the next section.
Activate Tare Mode
To activate either Automatic or Manual tare Mode.
When HHC is in Measurement Mode
Press TARE key

Press PORT SELECT key repeatedly to select Channel 1,

Channel 2, or both active pressure modules
Press ENTER
Selected modules are now in Tare Mode. The value displayed is

equal to the actual applied pressure less the tare value.
When TARE function is active the word “Tare” will be displayed

below the engineering units.