Cal password – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 67

Advertising
background image

65

f.

Press left and RIGHT ARROWs to position cursor

beneath desired character

g. Press UP and DOWN ARROWs to increment numeric

characters

h. When the password is correct press ENTER
i. “Calibrate” screen appears

Note: If the incorrect password is entered, the HHC display will

show a message “Bad Cal PW”, and the user will be asked to

Retry or Abort.

j. Press DOWN or UP ARROW key to highlight “System”

k. Press ENTER

l. “System Cal Ch1” screen appears

m. Press DOWN or UP ARROW key to highlight

“Begin Base Cal”

n. Press ENTER

10.

The HHC Base Unit will now begin prompting the operator to

apply precision voltage values to specific terminals, data col-

lection will take a few seconds then the HHC will prompt the

operator to move to the next step in the following sequence :

a. Apply 0.00000 volts to calibration module press ENTER
b. Apply 1.00000 volts to calibration module press ENTER

WARNING: Applying more than 1.25 volts to the calibration

module input could cause damage to the input of the HHC

11. When Prompted, set the voltage standard to 0.000 volts then
12.

POWER DOWN THE VOLTAGE STANDARD AND DISCON-

NECT THE VOLTAGE STANDARD FROM THE CALIBRA-

TION MODULE

13.

Connect the negative lead of the voltage standard capable of

delivering up to 10.000v DC to the black ringed “COM” input

jack

14.

Connect the positive lead of the voltage standard to the red

ringed “0-33Vdc In” jack

15.

Power on a voltage standard and allow it to warm up to its

specified accuracy as defined by the manufacturer.

16.

The HHC Base Unit will prompt the operator to apply preci-

sion voltage values to specific terminals, data collection will

take a few seconds then the HHC will prompt the operator to

move to the next step in the following sequence :

a.

Apply 0.000 volts to the voltage input jacks press ENTER

b.

Apply 10.000 volts to the voltage input jacks press

ENTER

17.

If using a different voltage standard for 0-30 volt calibration:

POWER DOWN THE VOLTAGE STANDARD AND DISCON-

NECT INPUT TERMINALS

18.

Connect the negative lead of the voltage standard capable of

delivering up to 30.000v DC to the black ringed “COM” input

jack

19.

Connect the positive lead of the voltage standard to the red

ringed “0-33Vdc In” jack

Cal Password

00000

^

Enter to accept

Esc to cancel

SECTION 15 n.

CALIBRATE BASE UNIT CONT.